Imprtant: This software and firmare update can be be utilized on instruments with firmaware versions 01.61 and 01.62. 
If the firmware version on the instrument are not those listed above, do not perform the update and contact Promega tech support for assistance.
To perform the update follow these steps:
1. Copy the executable file (*.exe) provided to the root of the USB stick and insert into the USB drive on the instrument. 
2. Press the Ok button when asked to update software. 
3. Remove executable file from the USB stick after process is complete. 
4. Copy the firmware files (af.ldf, ff.ldf and fc.ldf) provided to the root of the USB stick and insert into the USB drive on the instrument. 
5. Press the Ok button when asked to update firmware. 
6. Remove the firmware files from the USB stick after process is complete. 